Oil Analysis Automatic Distillation Analyzer Chemical Lab Solutions

The Automatic Distillation Analyzer is designed to perform optimal distillation analyses of gasolines, fuels, oils, solvents, aromatics, napthas, kerosenes, hydrocarbons, and other volatile products to ensure conformity to rigid quality control standards. The analyzer automatically performs tests, processes results, and produces standard reports according to ASTM, ISO, and related specifications.

Features

  • Conforms to ASTM D86, D285, D4737 and related international specifications
  • Pt-100 RTD probe with automatic temperature calibration system (°C or °F)
  • Windows®-based software package for PC control with LIMS export capabilities
  • Automatic determination of initial boiling point (IBP), final boiling point (FPB), dry point and barometric and residue corrections
  • Diagnostic system continuously ensures proper unit performance and user safety
  • Automatic temperature and volume calibration
  • Programmable distillation rate (2-15mL/min)
  • Ready for distillation groups 1 - 4
  • Networking for up to 32 units
  • Powerful CFC-free cooling and heating system
  • Receiver chamber heating system up to 60°C
  • Precision level follower system with optical meniscus detector
  • Integrated automatic fire extinguishing system with manual operation override

Specifications

Conforms to the specifications of: ASTM D86, D285, D850, D1078, D4737; D189 Section 10; DIN 51751; ISO 3405; IP 123; IP 195; JIS K0066; JIS K2254; NF M 07-002

Temperature Distillation Range: 0 to 450°C (±0.1°C accuracy)

Condenser: -5 to 60°C (±0.1°C accuracy); closed loop system

Receiver Chamber: 0 to 60°C (±0.1°C accuracy)

Distillation Rate: 2 to 15mL/min in 0.1mL increments, user selectable

Receiver Volume: 0 to 100mL (±0.01mL accuracy) by photoelectric infrared detection of meniscus by level following system utilizing a precision stepper motor and a special calibrated glass receiver; automatic calibration of evaporated loss volume and automatic volume calibration system ensures highest accuracy

Barometic Pressure: Automatic barometric correction utility with automatic sensor, range 550 to 900 mm Hg (±1 mm Hg accuracy)

Dry Point Detection: Automatic dry point detection board is included with standard equipment and only requires a dry point sensor, 200mL flask and PTFE plug for ASTM D850 and D1078 tests.

Environment: Operates at 0 to 25°C (113°F)